Siemens AG RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems

CVE-2018-4846
CRITICAL 9.8

At a glance

Source
NVD
EPSS
EPSS 1.76 % · percentile 78 ()
State
PUBLISHED
Published
Updated
Assigner
siemens
NVD status
Modified

Description

A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.

Weaknesses (CWE)

  • CWE-798 — Use of Hard-coded Credentials

References

Scores

VersionScoreVectorScored by
CVSS 3.0CRITICAL 9.8CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HNVD · nvd@nist.gov (Primary)
CVSS 2.0HIGH 10.0AV:N/AC:L/Au:N/C:C/I:C/A:CNVD · nvd@nist.gov (Primary)

European references

CERT-FR advisories

Affected configurations (NVD CPE)

Affected products

ProductVersions
Siemens AG · RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systemsaffected RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions < V3.3 _with_Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions >= V3.0 _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems V2.4.X _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions =< V2.3 _with_ Siemens Healthineers Informatics products RAPIDPoint 400 systems All versions_with_ Siemens Healthineers Informatics products

Sources

SourceLicence
CVE List (CVEProject cvelistV5)CVE Program Terms of Use
EPSS (FIRST.org)FIRST — attribution requested
ENISA EUVDENISA — reproduction authorised, source acknowledged
CERT-FR (ANSSI)Licence Ouverte / Etalab 2.0
NVD (NIST) via fkie-cad/nvd-json-data-feedsPublic domain — this product uses data from the NVD API but is not endorsed or certified by the NVD